• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH, INC. KANGAROO EPUMP ENTERAL FEED; PUMP, INFUSION, ENTERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH, INC. KANGAROO EPUMP ENTERAL FEED; PUMP, INFUSION, ENTERAL Back to Search Results
Model Number 382400
Device Problems Improper Flow or Infusion (2954); Program or Algorithm Execution Problem (4033)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/15/2022
Event Type  malfunction  
Event Description
Upon inspection of the kangaroo pump, it was noted that the programmed continuous feeds were set at a rate of 55cc/hr.The free water fluids programmed into the pump were set for 150cc every 4 hours.The hourly pump history did correctly reflect the continuous infusion but the flush history revealed that no free water flushes were administered automatically by the pump.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO EPUMP ENTERAL FEED
Type of Device
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
CARDINAL HEALTH, INC.
3651 birchwood drive
waukegan IL 60085
MDR Report Key15484800
MDR Text Key300564377
Report Number15484800
Device Sequence Number1
Product Code LZH
UDI-Device Identifier10884521006614
UDI-Public(01)10884521006614
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Risk Manager
Type of Report Initial
Report Date 09/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number382400
Device Catalogue Number382400
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA09/15/2022
Event Location Hospital
Date Report to Manufacturer09/26/2022
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/26/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age31390 DA
Patient SexMale
Patient Weight75 KG
-
-