• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THORATEC SWITZERLAND GMBH CENTRIMAG 2ND GENERATION PRIMARY CONSOLE WITH EM-TEC ADULT FLOW PROBE, OUS; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THORATEC SWITZERLAND GMBH CENTRIMAG 2ND GENERATION PRIMARY CONSOLE WITH EM-TEC ADULT FLOW PROBE, OUS; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number L201-90401
Device Problems Mechanical Problem (1384); Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/20/2022
Event Type  malfunction  
Event Description
It was reported that the centrimag console was defective.
 
Manufacturer Narrative
No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is completed.Reporter address line 1: (b)(6).
 
Manufacturer Narrative
Manufacturer's investigation conclusion: the reported event of an s3 error correlating to an error in internal battery operation was unable to be confirmed.A review of the downloaded log file contained data spanning approximately 13 days (2sep22 ¿ 3sep22, 8sep22, 19sep22, 2oct22, 5oct22, 8oct22 ¿ 10oct22, 18oct22, 4nov22, 9nov22, 18jan23 per timestamp).There were no s3 alarms or notable alarms active in the log file.Pump operation was not affected.The centrimag 2nd generation primary console (s/n: l02080-0007) was returned to the european distribution center (edc) for evaluation.The centrimag underwent functional testing and operated as intended.Atypical alarms were unable to be reproduced throughout testing.The internal battery was observed to have been past its expiration date and was replaced.It is unable to be determined if the battery was in use while expired.Due to the age of the console, the unit was scrapped and not made available in the loaner pool.A root cause of the reported event was unable to be conclusively determined through this analysis; however, the battery being past expiration may have contributed to the event.Review of the device history record for centrimag console, serial number l02080-0007, showed the device was manufactured in accordance with manufacturing and qa specifications.The 2nd generation centrimag system operating manual (rev.L) section 3 "warnings and precautions" warns "one additional 2nd generation centrimag primary console, motor and flow probe are required as backup system in the immediate vicinity of each patient whenever the centrimag or pedivas blood pump is used.The backup console must be connected to the backup motor and to the backup flow probe, have a battery charge sufficient for at least one hour of operation, be connected to ac power (except during transport) and be immediately available should the main console, motor or flow probe experience a malfunction." the 2nd generation centrimag system operating manual (rev.L, ¿table 15: console maintenance schedule¿) instructs users to have the centrimag console¿s internal battery replaced every two years.Users are instructed to contact abbott for assistance in replacing the battery, as users may not replace the internal battery without proper training or assistance.The 2nd generation centrimag system operating manual (rev.L) section 11.1-"appendix i ¿ console alarms and alerts" contains a list of console alarms and alerts, including s3 alarms, as well as appropriate operator response to these events.No further information was provided.The manufacturer is closing the file on this event.
 
Event Description
The centrimag alarmed with an s3 alarm that correlated to error in battery operation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIMAG 2ND GENERATION PRIMARY CONSOLE WITH EM-TEC ADULT FLOW PROBE, OUS
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ  CH-8005
Manufacturer (Section G)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ   CH-8005
Manufacturer Contact
bob fryc
6035 stoneridge drive
pleasanton, CA 94588
7818528204
MDR Report Key15486647
MDR Text Key306350689
Report Number3003306248-2022-12520
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K020271
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberL201-90401
Device Lot Number5467990
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/20/2022
Initial Date FDA Received09/26/2022
Supplement Dates Manufacturer Received02/22/2023
Supplement Dates FDA Received02/23/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/18/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-