Manufacturer's investigation conclusion: the reported event of an s3 error correlating to an error in internal battery operation was unable to be confirmed.A review of the downloaded log file contained data spanning approximately 13 days (2sep22 ¿ 3sep22, 8sep22, 19sep22, 2oct22, 5oct22, 8oct22 ¿ 10oct22, 18oct22, 4nov22, 9nov22, 18jan23 per timestamp).There were no s3 alarms or notable alarms active in the log file.Pump operation was not affected.The centrimag 2nd generation primary console (s/n: l02080-0007) was returned to the european distribution center (edc) for evaluation.The centrimag underwent functional testing and operated as intended.Atypical alarms were unable to be reproduced throughout testing.The internal battery was observed to have been past its expiration date and was replaced.It is unable to be determined if the battery was in use while expired.Due to the age of the console, the unit was scrapped and not made available in the loaner pool.A root cause of the reported event was unable to be conclusively determined through this analysis; however, the battery being past expiration may have contributed to the event.Review of the device history record for centrimag console, serial number l02080-0007, showed the device was manufactured in accordance with manufacturing and qa specifications.The 2nd generation centrimag system operating manual (rev.L) section 3 "warnings and precautions" warns "one additional 2nd generation centrimag primary console, motor and flow probe are required as backup system in the immediate vicinity of each patient whenever the centrimag or pedivas blood pump is used.The backup console must be connected to the backup motor and to the backup flow probe, have a battery charge sufficient for at least one hour of operation, be connected to ac power (except during transport) and be immediately available should the main console, motor or flow probe experience a malfunction." the 2nd generation centrimag system operating manual (rev.L, ¿table 15: console maintenance schedule¿) instructs users to have the centrimag console¿s internal battery replaced every two years.Users are instructed to contact abbott for assistance in replacing the battery, as users may not replace the internal battery without proper training or assistance.The 2nd generation centrimag system operating manual (rev.L) section 11.1-"appendix i ¿ console alarms and alerts" contains a list of console alarms and alerts, including s3 alarms, as well as appropriate operator response to these events.No further information was provided.The manufacturer is closing the file on this event.
|