• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT SBP DR INDUCTIVE PACEMAKER, JPD; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT SBP DR INDUCTIVE PACEMAKER, JPD; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM2152
Device Problems Failure to Interrogate (1332); No Pacing (3268)
Patient Problems Fatigue (1849); Syncope/Fainting (4411)
Event Date 09/09/2022
Event Type  Injury  
Event Description
It was reported that the patient had syncope and fatigue.The device could not be interrogated, did not respond to a magnet, and was not pacing.The device was explanted and replaced to resolve the event.The patient was in stable condition following the procedure.
 
Manufacturer Narrative
The reported events of no pacing, inability to interrogate, and no magnet response were confirmed.The device was received with no telemetry communication and no output.Visual inspection of the header attachment area detected an anomaly between the pre-molded header and titanium case.The device was cut open to enable further testing and the battery was found depleted.Feedthrough leak test was performed, indicating a device hermeticity breach.This is consistent with feedthrough damage as a result of fluid intrusion between the header and case, and subsequent fluid ingress to internal electronics.Hybrid circuitry was tested by connecting to an external power source.Test results indicated elevated current drain, consistent with moisture damage, depleting the battery and resulting in the reported events.A manufacturing process anomaly may have occurred, which resulted in the header bonding anomaly.The device is included in the zenex, assurity, endurity laser adhesion preparation advisory issued by abbott on 20 july 2022 for a subset of devices distributed and implanted outside of the united states.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SBP DR INDUCTIVE PACEMAKER, JPD
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15486896
MDR Text Key300640272
Report Number2017865-2022-39333
Device Sequence Number1
Product Code NVZ
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P880086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/26/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2021
Device Model NumberPM2152
Device Lot NumberP000093309
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/15/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/22/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-