Model Number 71992-01 |
Device Problem
High Readings (2459)
|
Patient Problems
Hyperglycemia (1905); Muscle Weakness (1967); Diaphoresis (2452)
|
Event Date 09/10/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
The product has been requested back for an investigation.A follow-up report will be submitted once additional information is obtained.The device mfg date is unknown.The date entered is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A high reading issue was reported with the adc device.A caller reported receiving an unspecified high sensor scan when compared to a built-in meter.The customer experienced paleness, weakness, and sweating, and was provided humalog insulin (dose unknown) based on the sensor reading.Consequently, the customer became hypoglycemic and required treatment of oral glucose from their mother.No further information was reported.There was no report of death or permanent injury associated with this event.
|
|
Manufacturer Narrative
|
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.No product has been returned.Extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Dhrs for the fs libre sensor and fs libre sensor kit were reviewed and the dhrs showed the fs libre sensor and sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A high reading issue was reported with the adc device.A caller reported receiving an unspecified high sensor scan when compared to a built-in meter.The customer experienced paleness, weakness, and sweating, and was provided humalog insulin (dose unknown) based on the sensor reading.Consequently, the customer became hypoglycemic and required treatment of oral glucose from their mother.No further information was reported.There was no report of death or permanent injury associated with this event.
|
|
Search Alerts/Recalls
|