|
Model Number 45007 |
Device Problems
Display or Visual Feedback Problem (1184); Difficult to Remove (1528)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/31/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that the catheter was difficult to remove.A 2.1mm jetstream xc catheter was selected for use in an atherectomy procedure.The target lesion was located in the leg and was moderately calcified.At the completion of the jetstream therapy, a bubl error occurred.The technician attempted to remove the catheter over the guidewire, however, it was extremely difficult to remove.The catheter was ultimately able to be removed from the guidewire and patient.Angioplasty was performed to complete the procedure.There were no patient complications.
|
|
Manufacturer Narrative
|
Device eval by mfr: the device was visually and microscopically examined for any damage.Visual examination revealed multiple minor buckling along the sheath.The device showed damage to the infusion sheath 14.5cm from the strain relief.Functional testing was performed and during testing the damaged section of the sheath would bulge.Testing was stopped to prevent potentially bursting the sheath.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.
|
|
Event Description
|
It was reported that the catheter was difficult to remove.A 2.1mm jetstream xc catheter was selected for use in an atherectomy procedure.The target lesion was located in the leg and was moderately calcified.At the completion of the jetstream therapy, a bubl error occurred.The technician attempted to remove the catheter over the guidewire, however, it was extremely difficult to remove.The catheter was ultimately able to be removed from the guidewire and patient.Angioplasty was performed to complete the procedure.There were no patient complications.
|
|
Search Alerts/Recalls
|
|
|