Catalog Number CA1300/0K1/000JP
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/01/2020
Event Type  malfunction  
Event Description
It was reported that during a pre-use check, air leakage was observed in the product.No patient injury was reported.
 
Manufacturer Narrative
A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample was received for evaluation.Visual and functional testing were performed.Visual observation of the anesthesia circuit, no abnormality such as damage or deformation was found.A leak test was conducted of the anesthesia circuit, but no leak was detected, and it was confirmed that the test complied with the standard.The reported event was not confirmed.Udi is unknown.Operator of device is unknown.Pma/510k is unknown.No information has been provided to date.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# 617147.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX ANESTHESIA BREATHING CIRCUIT
Type of Device
CIRCUIT, BREATHING W CONNECTOR, ADAPTOR, Y PIECE
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
2jo kudaru, kawaharamachidori,
minneapolis, MN 55442
MDR Report Key15492016
MDR Text Key306171577
Report Number3012307300-2022-20357
Device Sequence Number1
Product Code CAI
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCA1300/0K1/000JP
Device Lot Number200121
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/10/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/13/2022
Initial Date FDA Received09/27/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-