U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1458Q/86
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Cardiac Perforation (2513)
Event Date 09/07/2022
Event Type  Injury  
Event Description
It was reported that the patient presented to the hospital for a system implant procedure on (b)(6) 2022.While attempting to implant the left ventricular (lv) lead, it was noted that there was cardiac perforation due to the lead.The patient felt sick and the whole system was removed without further incident in the same procedure.There were no adverse consequences to the patient.
 
Manufacturer Narrative
The reported event was cardiac perforation.As received, a complete lead was returned in one piece for analysis.S-curve height of the lead was measured within specification.Visual and x-ray inspection of the lead did not find any anomalies with the exception of damage consistent with that occurring at the time of the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15492133
MDR Text Key300636054
Report Number2017865-2022-39419
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeIN
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/86
Device Catalogue Number1458Q-86
Device Lot NumberA000121898
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/07/2022
Initial Date FDA Received09/27/2022
Supplement Dates Manufacturer Received11/03/2022
Supplement Dates FDA Received11/15/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/26/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
TENDRIL; TENDRIL
Patient Outcome(s) Required Intervention;
-
-