A complaint was received reporting: "as reported by a field clinical specialist (fcs), during valve deployment of a 29 mm sapien 3 valve in the native aortic valve via transfemoral approach, the atrion ql38 inflation device cracked with approximately 3cc still in the inflation device.A 20cc syringe was connected to the delivery system to complete the deployment.A second atrion ql38 was open and used to insure the valve was fully inflated.Currently the device is with r&d for ct scanning, and we plan to conduct similar diagnostics as the previous investigation.".
|