Model Number GK370P
Device Problem Arcing of Electrodes (2289)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/30/2021
Event Type  malfunction  
Manufacturer Narrative
The product was sent to the production department for the subject matter expert analysis.The results are as follows: we have no indication for a production related error.We detected that the tip of the shaft was damaged.It could be possible that liquid had gotten between the insulation tube and the hook electrode during the operation.The liquid then conducted the current.This is known as capillary action.It is important that the product does not run in continuous load.For safe handling and preparation of the product, please note the points in the instructions for use (ifu).Batch history review: due to the fact that no lot number was provided, a review of the device history records for the complained device is not possible.The review of risk assessment revealed that the overall risk level ((3)5 severity x (2)5 probability of occurrence) according to din en iso 14971 is still acceptable.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material, manufacturing or design-related failure.Based on the investigation results a capa is not required.
 
Event Description
It was reported that there was an issue with gk370p - shaft only f/modular monopol.Electr.According to the complaint description, sparks came out near hook electrode.During use with vio30w, a spark came out from the root near the fitting of the hook electrode.According to the test results from a hospital scan, only the l-shaped non-insulated part at the tip makes noise and is insulated, but that is exactly where the sparks came out.Observation with a loupe showed that the hook was scratched, but no peeling of the insulation was found, and the deformation of the tip of the insulation part of the handle was caused by this spark.(inspected before use every time) there was no described patient harm.Additional information was not provided nor available.Additional patient information is not available.The malfunction is filed under aag reference (b)(4).Associated medwatch reports: gk370p (9610612-2022-00282) (b)(4).Involved components: gk373r - inner tube w/handle for gk372r (b)(4).Gk384r - ceramic electrode tip l-hk f/gk372r (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHAFT ONLY F/MODULAR MONOPOL.ELECTR
Type of Device
REUSABLE INSTRUMENTS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
christian von der grün
po box 40
tuttlingen, 78501
GM   78501
MDR Report Key15498136
MDR Text Key305155355
Report Number9610612-2022-00282
Device Sequence Number1
Product Code GEI
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K970541
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberGK370P
Device Catalogue NumberGK370P
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/30/2022
Initial Date FDA Received09/28/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
GK373R - INNER TUBE W/HANDLE FOR GK372R.; GK384R - CERAMIC ELECTRODE TIP L-HK F/GK372R.
-
-