MAUDE Adverse Event Report: CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM (VIPERWIRE GUIDE WIRE); PERIPHERAL ATHERECTOMY DEVICE (GUIDE WIRE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM (VIPERWIRE GUIDE WIRE); PERIPHERAL ATHERECTOMY DEVICE (GUIDE WIRE) Back to Search Results
Model Number VPR-GW-FLEX14
Device Problem Material Separation (1562)
Patient Problem Foreign Body In Patient (2687)
Event Date 09/19/2022
Event Type  Injury  
Event Description
A diamondback 360 peripheral orbital atherectomy device (oad) was used to treat the superior femoral artery and anterior tibial artery (at).Post treatment, the viperwire advance guide wire became stuck in the oad.The guidewire and oad were removed.Visual examination revealed the tip of the flex tip wire was not intact and remained in the at.No attempt was made to remove the fractured component as the branch was at complete total occlusion.The fractured component did not diminish flow and there was no concern to leave the component in vivo.The patient was stable.
 
Manufacturer Narrative
The material inspection report for this guidewire lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.The results of the investigation are inconclusive since the reported device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.Csi id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM (VIPERWIRE GUIDE WIRE)
Type of Device
PERIPHERAL ATHERECTOMY DEVICE (GUIDE WIRE)
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer (Section G)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer Contact
tonia moskalets
1225 old hwy 8 nw
st. paul, MN 55112
MDR Report Key15498572
MDR Text Key300707211
Report Number3004742232-2022-00232
Device Sequence Number1
Product Code MCW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190634
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 09/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/28/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2022
Device Model NumberVPR-GW-FLEX14
Device Catalogue Number7-10041-03
Device Lot Number345878
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/19/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/22/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age74 YR
Patient SexMale
-
-