H4: the device was manufactured in january 2022.Device history record (dhr) provided do not include packaging or labeling operations, for this reason a final manufacture date is not available.This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The dhr was reviewed and showed the product met all specifications upon release.Likely manufacture practices did not cause the reported failure.The op seq 70 (continuity test) and 90 (qa inspection) shows all finished units passed acceptance criteria provided in manufacturing procedure mpn-34-042.Based on the investigation, the reported complaint was likely due to mishandling during use.However, the exact root cause of the issue could not be conclusively specified.The instruction for use (ifu-99-0431_dg) states "keep the distal tip of any electrode, probe, laser fiber, or other ancillary device in the field of view at all times when active;" "should a higher power setting than expected be required, gradually increase the power.Note the tissue effect after each increase.If there is little or no change in the tissue effect, stop.Recheck all cables, connections, patient contacts, and irrigation fluid before proceeding.Recheck power setting;" "use only nonconductive irrigation solutions.Do not use saline or lactated ringers;" " refer to the appropriate operating and maintenance manual for the endoscope with which the flexible electrode is to be used for additional cautions and instructions" (warnings, page 2).
|