ZOLL MEDICAL CORPORATION ASSY,MAIH, 3.0 DIA, W/SWITCH, 100-120VONLY, PACKAG; INTERNAL PADDLE Back to Search Results
Model Number 1011-0139-04
Device Problem Failure to Deliver Shock/Stimulation (1133)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
Complainant alleged that during functional testing, the associated device failed to discharge using these internal handles.Complainant did not indicate that there was any patient involvement in the reported malfunction.
 
Manufacturer Narrative
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
 
Manufacturer Narrative
The internal handles will not be returned to zoll medical corporation.The biomed reported to zoll medical corporation that an internal inspection found a cut in the cable insulation showing internal wiring.Due to the age of these internal handles (47 months), they have most likely met their useful life and exceeded the maximum of 50 autoclave/sterilization cycles.The customer will remove this set of internal handles from service and will order a replacement set.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSY,MAIH, 3.0 DIA, W/SWITCH, 100-120VONLY, PACKAG
Type of Device
INTERNAL PADDLE
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key15501442
MDR Text Key300769403
Report Number1220908-2022-03681
Device Sequence Number1
Product Code LDD
UDI-Device Identifier00847946022358
UDI-Public00847946022358
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number1011-0139-04
Device Catalogue Number8011-0139-04
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received09/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/01/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-