Device Problem Firing Problem (4011)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/27/2022
Event Type  malfunction  
Event Description
Procedure performed: laparoscopic cholecystectomy.Event description: the rep was not present during the case.During the case the surgeon used the device to clip the cystic artery.The surgeon initially believed that the artery was successfully clipped but when they went to transect the artery the vessel still bled.The surgeon believed that the clip applied was not tight enough on the vessel.It is unknown if this was the first clip from the device or a subsequent clip.It is unknown which trocar the device was used through.A new ca500 clip applier was used to clip the artery again and was able to successfully stop the bleeding from the vessel.The bleeding was considered minor and no blood transfusion was necessary.No patient injury.Product is available for return.Additional information received on 01jul2022 from [name], applied medical account manager i.The issue occurred on the third clip firing.The clip was fully loaded into the jaws before firing.The trigger was squeezed plastic to plastic.The vessel was fully skeletonized prior to clip application.The surgeon did not use the applier to skeletonize the tissue.The surgeon stated the tip did close but it did not fully close the entire clip upon plastic-to-plastic squeeze.Information from medwatch report number (b)(4) received via email on 11jul2022: provider was performing a cholecystectomy on a patient, and was attempting to clamp the cystic artery when the clip applier failed, and the arterial clip was not clamped appropriately.The provider had to request a new device in order to perform the clamping of the clip correctly.Additional information received on 12jul2022 via email from [name], account manager.That clip was apparent of the lap chole kit and was purchased by account (b)(6).Additional information received on 12jul2022 via email from [name], account manager: lot number confirmed to be 1450652.Additional information received on 15jul2022 via email from [name], interim quality director."the patient suffered no injury.The vessel was completely skeletonized using a maryland dissector before applying the clips.The tips of the clips were approximated but it was later realized (after firing additional clips) that the apex was not closed appropriately.No dissection was performed with the clip applier and the instrument was only used for the clips as i generally use it (clamping down firmly, plastic to plastic).When bleeding occurred i placed a new clip and made sure to clamp firmly, but bleeding continued.Nonfunctioning clips were removed, the bleeding was controlled with the maryland, and a new clip applier was obtained.There was not any significant blood loss, just more than i usually have for a lap chole." additional information received on 09sep2022 via email from [name], director of surgical services, [name of institution].There was only one incident at [name of institution].Patient status: no patient injury.Intervention: bleeding was controlled with the maryland.Used new ca500 device to complete the case.
 
Manufacturer Narrative
The event unit was returned to applied medical for evaluation.Functional testing was performed on the returned unit, where the remaining clips closed properly when fired.The complainant¿s experience could not be replicated or confirmed as the device passed functional testing.Applied medical has reviewed the details surrounding the event and is unable to determine the cause of the reported event or confirm that a product malfunction occurred.The probability and criticality of harm resulting from this failure have been evaluated and were found to be at an acceptable level.The event is not reportable as it is unlikely to cause or contribute to death or serious injury.This report is to follow-up medwatch # (b)(4).Medwatch report #2027111-2022-00660 (documented under complaint # (b)(4)) was initially submitted to follow-up medwatch # (b)(4).However, upon further evaluation of complaint # (b)(4), it was determined that this incident had already been documented under complaint # (b)(4).As complaint # (b)(4) and medwatch #2027111-2022-00660 are being voided, the current medwatch will follow-up medwatch # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CA500, EPIX UNIVERSAL CLIP APPLIER 3/BX
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
Manufacturer Contact
aaron fulcher
22872 avenida empresa
rancho santa margarita, CA 92688
9497135765
MDR Report Key15503110
MDR Text Key306320658
Report Number2027111-2022-00745
Device Sequence Number1
Product Code FZP
UDI-Device Identifier00607915125318
UDI-Public(01)00607915125318(17)250411(30)01(10)1450652
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K011236
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCA500
Device Catalogue Number101474072
Device Lot Number1450652
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/18/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/11/2022
Initial Date FDA Received09/29/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/12/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
5 MM ADV TROCAR
-
-