Model Number AGFN-756
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Heart Block (4444)
Event Date 09/07/2022
Event Type  Injury  
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
Clinical information: crd_1036 - regent (b)(6) pmcf, patient site id: (b)(6).It was reported that on (b)(6) 2022, a 21mm sjm regent heart valve with flex cuff was successfully implanted during an aortic valve replacement.A mitral valve replacement also was performed during the procedure.An abbott sizer set was used during procedure.There were no intraprocedural adverse events.Post procedure but prior to discharge from the hospital, on (b)(6) 2022, an electrocardiogram (ecg) showed a third degree atrioventricular (av) block.On (b)(6) 2022, a permanent pacemaker was successfully implanted.This event prolonged the patient's hospitalization.The cause of the av heart block was unknown but suspected to be possibly related to the procedure itself.
 
Event Description
N/a.
 
Manufacturer Narrative
An event of third degree atrioventricular (av) block was reported.A returned device assessment could not be performed as the device remains implanted and was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REGENT HEART VALVE WITH FLEX CUFF
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC. REG#2648612
20 b st caguas west park
caguas 00725
*   00725
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key15506108
MDR Text Key300875154
Report Number2135147-2022-01410
Device Sequence Number1
Product Code LWQ
UDI-Device Identifier05414734005975
UDI-Public05414734005975
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Study,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/29/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberAGFN-756
Device Catalogue Number21AGFN-756
Device Lot Number8311687
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/26/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age55 YR
Patient SexMale
Patient Weight54 KG
-
-