Model Number CI-1400-01 |
Device Problems
Mechanical Problem (1384); Battery Problem (2885); Output Problem (3005); Intermittent Communication Failure (4038)
|
Patient Problem
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 09/15/2022 |
Event Type
malfunction
|
Event Description
|
The recipient reportedly experienced no sound and short battery life.Device testing revealed results within normal limits.External equipment was exchanged and programming adjustments were made and the issues improved slightly, however, the recipient proceeded with revision surgery.The recipient's device was explanted.The recipient was reimplanted with another cochlear device.
|
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed silicone damage on the top cover of the device.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock could not be obtained at any spacing.An electrical test could not be performed due to no lock.The device passed some of the electrical tests performed.The device passed the mechanical test performed.This device had moisture that exceeded the residual gas analysis (rga) test limit.Based on an assessment of the rga data, it is determined that this device was non-hermetic.Corrective actions were implemented.This version of the hires 90k device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Search Alerts/Recalls
|