KARL STORZ SE & CO. KG CUTTING LOOP, BIPOLAR; BIPOLAR SINGLE USE CUTTING LOOP Back to Search Results
Model Number 27040GP130-S
Device Problems Material Separation (1562); Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/07/2022
Event Type  malfunction  
Event Description
The manufacturer's complaint reference number for this case is (b)(4).Se edwin carty reported that the doctor had 2 bipolar loops break during separate procedures.The loops completely came off the tip.No patient harmed during the procedure.Both loops from the same lot.First complaint is (b)(4).
 
Manufacturer Narrative
The affected product will not be returned, no evaluation results will be obtained.Based on the information provided by the facility that the tip came off the loop in the middle of a procedure, mechanical damage is a probable cause.The ifu 97000002usca 02-2019, states the product should not be manipulated or bent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUTTING LOOP, BIPOLAR
Type of Device
BIPOLAR SINGLE USE CUTTING LOOP
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
78532
tuttlingen, gm
Manufacturer (Section G)
KARL STORZ SE & CO.KG
dr.-karl-storz -strasse 34
78532
tuttlingen, gm
Manufacturer Contact
anja fair
2151 e. grand avenue
el segundo, CA 90245
4242188247
MDR Report Key15508749
MDR Text Key306353533
Report Number9610617-2022-00273
Device Sequence Number1
Product Code FAS
UDI-Device Identifier04048551393864
UDI-Public4048551393864
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K061541
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number27040GP130-S
Device Catalogue Number27040GP130-S
Device Lot Number37CJ1218
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/13/2022
Initial Date FDA Received09/29/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age80 YR
Patient SexMale
Patient Weight88 KG
-
-