Model Number G148
Device Problems Premature Discharge of Battery (1057); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/05/2022
Event Type  Injury  
Manufacturer Narrative
Additional information was requested.The investigation will be updated should further information be provided.
 
Event Description
It was reported that this device recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.A request was made to have data from this device analyzed.Data analysis showed the battery appeared to be depleting more quickly than expected.Device replacement was recommended.This device remains in service.No adverse patient effects were reported.
 
Event Description
It was reported that this device recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.The device emitted beeping tones associated to the code 1003.A request was made to have data from this device analyzed.Data analysis showed the battery appeared to be depleting more quickly than expected.Device replacement was recommended.This device was replaced and is expected to be returned for analysis.No additional adverse patient effects were reported.
 
Manufacturer Narrative
Supplemental: at this time, the product has not been returned.If the product is returned, analysis will be performed, and this report would be updated should information be provided at that time.Initial: additional information was requested.The investigation will be updated should further information be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN X4 CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key15510719
MDR Text Key302267244
Report Number2124215-2022-38753
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534591
UDI-Public00802526534591
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
P010012/S341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/30/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/23/2020
Device Model NumberG148
Device Catalogue NumberG148
Device Lot Number189134
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/28/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/04/2018
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-