Model Number G148 |
Device Problems
Premature Discharge of Battery (1057); Noise, Audible (3273)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/05/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
Additional information was requested.The investigation will be updated should further information be provided.
|
|
Event Description
|
It was reported that this device recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.A request was made to have data from this device analyzed.Data analysis showed the battery appeared to be depleting more quickly than expected.Device replacement was recommended.This device remains in service.No adverse patient effects were reported.
|
|
Event Description
|
It was reported that this device recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.The device emitted beeping tones associated to the code 1003.A request was made to have data from this device analyzed.Data analysis showed the battery appeared to be depleting more quickly than expected.Device replacement was recommended.This device was replaced and is expected to be returned for analysis.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Supplemental: at this time, the product has not been returned.If the product is returned, analysis will be performed, and this report would be updated should information be provided at that time.Initial: additional information was requested.The investigation will be updated should further information be provided.
|
|