The reported event of pacing issue was unable to be confirmed.No visible damage or abnormality was observed from catheter body, balloon, windings and returned syringe.Continuity testing was performed on the distal and proximal circuits and there were no open, intermittent, or short conditions observed.The balloon inflated clear and concentric with 1.3 cc air and the balloon remained inflated for 5 min.Without leakage.Further evaluation regarding related quality issues is under investigation.A device history record review was completed and documented that device met all specifications upon distribution.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
|