• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM; MITRAL VALVE REPAIR DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM; MITRAL VALVE REPAIR DEVICES Back to Search Results
Model Number N/A
Device Problems Unstable (1667); Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/11/2022
Event Type  malfunction  
Manufacturer Narrative
All available information was investigated, and the reported clip close inability was confirmed via returned device analysis.In addition, it was observed that the actuator mandrel was protruding from the proximal end, the release crimper was loose, and the collet was cracked.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents reported from this lot.Based on available information, the reported clip close inability was due to the cracked collet tine.The observed protruding actuator mandrel was due to the cracked collet tine.The investigation determined the loose release crimper and the cracked collet to be related to a potential product quality issue.The issue is being addressed per internal operating procedures.Abbott will continue to trend the performance of these devices.
 
Event Description
This is filed to report an unstable/loose release crimper found during device analysis.It was reported that the initial mitraclip procedure was performed on (b)(6) 2022, to treat functional mitral regurgitation (mr) with a grade of 4.During device preparation, the clip was tested by opening and closing the clip by rotating the arm positioner.The lock lever was opened to the blue line and locked after 60 degrees.When completely closing the clip, the clip was not able to close more than 20%.The clip was able to open but not fully close.Multiple attempts were made unsuccessfully and the clip was not used.There was no patient involvement.Another clip was used to complete the procedure with no issues.One clip total was implanted reducing the mr to <1.There was no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key15522766
MDR Text Key306331762
Report Number2135147-2022-01433
Device Sequence Number1
Product Code NKM
UDI-Device Identifier08717648231001
UDI-Public08717648231001
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 09/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/30/2023
Device Model NumberN/A
Device Catalogue NumberCDS0701-XTW
Device Lot Number20131R286
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/12/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/07/2022
Initial Date FDA Received10/01/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/31/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-