Device analysis: visual inspection of the returned device found the proximal filter sheathed, the distal filter sheathed, the articulating sheath ads relaxed, the hypotube kinked and the outer shaft kinked.X-ray inspection found the pull wire was damaged.Functional testing found before flushing, articulating sheath ads was unable to be flexed using knob #2 and the proximal filer was unable to be deployed.After flushing the proximal filter was unable to be deployed.Dissection and microscopic testing found the pull wire was detached/separated.A transversal cut was performed on the distal part of the outer shaft/sheath and no damage was found on the inner shaft.The proximal filter was found torn.
|
Reportable based on returned device analysis completed (b)(6) 2022.It was reported that the proximal filter failed to deploy.A sentinel cerebral protection system (cps) was used during a transcatheter aortic valve implantation (tavi) procedure.An introducer sheath was placed into the radial artery and the sentinel cps was advanced and positioned with the proximal filter into brachiocephalic artery.Following deployment of the proximal filter, lcc could not be achieved.The proximal filter was recaptured and repositioned, but it was unable to be released again.The sentinel cps was successfully retrieved and replaced with another sentinel cps to complete the procedure.No patient complications were reported.However returned device analysis found the proximal filter was found torn.
|