Model Number XS PST |
Device Problems
Display Difficult to Read (1181); Image Display Error/Artifact (1304)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/08/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Occupation is patient/consumer the investigation is ongoing.
|
|
Event Description
|
The customer complained of a display issue with coaguchek xs meter serial number (b)(4).Customer stated the meter display is faint and they cannot read the display clearly.They stated they can see some segments in the display but everything is faded.No misinterpretation of results was reported.
|
|
Manufacturer Narrative
|
The meter was returned for investigation.The meter did not power on during the investigation.The meter was disassembled for further investigation.The optical inspection of the printed circuit board revealed contamination due to residues of fluid ingress.The contamination on the battery contacts caused a power interruption.Therefore it is not possible to turn the meter on.The contamination can lead to the mentioned problem with faded or missing segments.The investigation determined the issue was due to improper handling or maintenance by the customer.
|
|