• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME MONITOR Back to Search Results
Model Number XS PST
Device Problems Display Difficult to Read (1181); Image Display Error/Artifact (1304)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/08/2022
Event Type  malfunction  
Manufacturer Narrative
Occupation is patient/consumer the investigation is ongoing.
 
Event Description
The customer complained of a display issue with coaguchek xs meter serial number (b)(4).Customer stated the meter display is faint and they cannot read the display clearly.They stated they can see some segments in the display but everything is faded.No misinterpretation of results was reported.
 
Manufacturer Narrative
The meter was returned for investigation.The meter did not power on during the investigation.The meter was disassembled for further investigation.The optical inspection of the printed circuit board revealed contamination due to residues of fluid ingress.The contamination on the battery contacts caused a power interruption.Therefore it is not possible to turn the meter on.The contamination can lead to the mentioned problem with faded or missing segments.The investigation determined the issue was due to improper handling or maintenance by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK ® XS SYSTEM
Type of Device
PROTHROMBIN TIME MONITOR
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15527565
MDR Text Key305995519
Report Number1823260-2022-03024
Device Sequence Number1
Product Code GJS
UDI-Device Identifier00365702125100
UDI-Public00365702125100
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/03/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberXS PST
Device Catalogue Number04837738001
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received10/06/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient SexMale
-
-