Model Number ER420 |
Device Problem
Firing Problem (4011)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/15/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during a laparoscopic cholecystectomy procedure, the surgeon went to place the device on the structure, pulled handle one time, the clip deployed and the handle broke, rendering the device unusable.A like device was used to complete the procedure with no patient harm.
|
|
Manufacturer Narrative
|
(b)(4).Batch # unk.A manufacturing record evaluation was performed for the finished device, er420 lot number x95f5e, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Manufacturer Narrative
|
(b)(4).Date sent: 11/8/2022 d4: batch # x9510e investigation summary the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the er420 device was received with the trigger broken.The device was disassembled to evaluate the condition of the internal components and the trigger was found broken, not allowing the clips to fed into the jaws.In addition, 19 clips were found remaining inside the clip track.The damaged on the device could be that the internal ribs were being gauged by the feeder link, causing the trigger to experienced additional force that could interfere with the firing of the device.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|
|
Search Alerts/Recalls
|
|
|