MAUDE Adverse Event Report: BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH¿ ELECTROPHYSIOLOGY CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH¿ ELECTROPHYSIOLOGY CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Catalog Number D133604IL
Device Problem Temperature Problem (3022)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/14/2022
Event Type  malfunction  
Manufacturer Narrative
The biosense webster, inc.Product analysis lab received the device on 29-aug-2022.The device evaluation was completed on 06-sep-2022.The thermocool® smart touch¿ electrophysiology catheter was returned to biosense webster (bwi) for evaluation.A visual inspection, temperature and impedance test of the returned device were performed following with bwi procedures.Visual analysis of the returned sample revealed the pebax was broken.The temperature and impedance test were performed, and no impedance was displayed due to an open circuit on the tip area.A manufacturing record evaluation was performed for the finished device 30672380m number, and no internal action was found during the review.The issue reported by the customer was confirmed.It should be noted that product failure is multifactorial.The instructions for use contain the following warning stated in the carto 3 system manual: if the rf generator does not display temperature, verify that the appropriate cable is plugged into the rf generator.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.This product issue will be addressed through bwi's quality system.The ¿suspected medical device¿ reported in section d of this report is not marketed in usa or approved by the fda.However, it is being reported as biosense webster considers this as a similar device to thermocool® smart touch¿ electrophysiology catheter(d133601) approved under pma # p030031/s053.This report is being submitted pursuant to the provisions of 21 cfr, part 4.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
 
Event Description
It was reported that a patient underwent an atrial tachycardia (at) procedure with a thermocool® smart touch¿ electrophysiology catheter and the biosense webster inc, (bwi) product analysis lab observed a broken pebax.Initially, it was reported that during the operation, there was no temperature displayed on the carto or generator.A second catheter was used to complete the operation.There was no adverse event reported on patient.The issue of no temperature was assessed as not mdr reportable.The potential that it could cause or contribute to a death or serious injury, or other significant adverse event, is remote.This event is being reported because the biosense webster inc, (bwi) product analysis lab received the device for evaluation and visual analysis of the returned sample revealed the pebax was broken.This finding was assessed as mdr reportable.Therefore, the awareness date for this reportable lab finding is 06-sep-2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SMART TOUCH¿ ELECTROPHYSIOLOGY CATHETER
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
michelleann garcia
31 technology dr
irvine, CA 92618
646591-798
MDR Report Key15544120
MDR Text Key306441402
Report Number2029046-2022-02404
Device Sequence Number1
Product Code LPB
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Type of Report Initial
Report Date 10/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/14/2022
Device Catalogue NumberD133604IL
Device Lot Number30672380M
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/29/2022
Initial Date Manufacturer Received 09/06/2022
Initial Date FDA Received10/05/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/15/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNK BRAND CATHETER.; UNK_CARTO 3.
-
-