• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. DERMATOME AN HANDPIECE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. DERMATOME AN HANDPIECE Back to Search Results
Model Number N/A
Device Problems Power Problem (3010); Intermittent Loss of Power (4016)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/03/2022
Event Type  malfunction  
Event Description
It was reported that during cleaning/inspection the device stalled.No harm or delay were reported as issue occurred when there was no patient involvement.No adverse event was reported as it relates to this event.Diligence is complete and no additional information is available.
 
Manufacturer Narrative
This incident has been recorded under (b)(4).Upon completion of investigation a final/supplemental will be submitted.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.The following sections were updated/corrected: b4, b5, d4, d9, g2, g3, g6, h2, h3, h4, h6, h10.Review of the most recent repair record determined the motor was running slow.The motor, swivel, eccentric shaft, bearing assembly, neck gland seal, and o-ring were replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No additional event information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DERMATOME AN HANDPIECE
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key15546338
MDR Text Key306495334
Report Number0001526350-2022-00965
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/05/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number88710100
Device Lot Number64310247
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/22/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/04/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-