Catalog Number 22445-31
Device Problem Difficult to Remove (1528)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/14/2022
Event Type  malfunction  
Event Description
It was reported that the procedure was to treat a femoropopliteal artery.A bare wire guide 315cm (not the barewire that came with the nav6 device) was advanced to the target lesion and then an emboshield nav6 filter was advanced as well to perform the procedure without issue.When attempting to remove the emboshield, resistance was met with the barewire 315cm.Both devices were removed altogether as a single unit.The procedure was successfully completed with a drug-filled balloon and stent.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation was unable to determine a conclusive cause for the reported difficulties.Based on the information reviewed, there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.The additional emboshield nav6 device referenced in event and concomitant medical products is filed under separate medwatch report number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BAREWIRE EMBOSHIELD FILTER DELIVERY WIRE
Type of Device
EMBOLIC PROTECTION SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15547036
MDR Text Key306510095
Report Number2024168-2022-10354
Device Sequence Number1
Product Code NTE
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K141678
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 10/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/05/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2023
Device Catalogue Number22445-31
Device Lot Number0120261
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/02/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
EMBOSHIELD NAV 6
-
-