• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL LEVEL 1 HOTLINE LOW FLOW SYSTEMS - HL-90; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL LEVEL 1 HOTLINE LOW FLOW SYSTEMS - HL-90; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-90
Device Problem Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/02/2022
Event Type  malfunction  
Event Description
It was reported that the light broke off.Additional info received via email on 30-sep-2022.Event date updated.No patient involved.
 
Manufacturer Narrative
Operator of device is unknown.No information has been provided to date.A product sample was received and is awaiting evaluation and investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
H10: device evaluation: the device was returned for investigation.A visual inspection and functional test were performed.Visual inspection found the device was received with broken leds (light emitting diode) and cracked enclosure under drain fitting.Worn line cord, front cover, missing swivel for the pole clamp, stained reservoir, interlock block screw broke off, faulty microswitch and cracked reservoir cover around all four corners.The customer stated problem was duplicated.Replaced pcb (printed circuit board), enclosure, drain fitting, tank cover, front cover, interlock block assay, microswitch, line cord, and pole clamp.The cause of the reported problem was user manipulation of the device (impact).A dhr (device history review) was not performed because the device is beyond a year from its manufacture date and there was no indication of a manufacturing defect during the investigation.Service history review identified this device has not been in for service previously.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE LOW FLOW SYSTEMS - HL-90
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
receiving dock
minneapolis, MN 55442
MDR Report Key15550963
MDR Text Key306813424
Report Number3012307300-2022-21817
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier50695085820909
UDI-Public50695085820909
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberHL-90
Device Catalogue NumberHL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/16/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/12/2022
Initial Date FDA Received10/06/2022
Supplement Dates Manufacturer Received11/23/2022
Supplement Dates FDA Received12/02/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/17/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-