DATEX-OHMEDA, INC. AVANCE CS2; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Suction Failure (4039)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/12/2022
Event Type  malfunction  
Event Description
It was reported that there was a malfunction resulting in loss of suction.There was no patient involvement.
 
Manufacturer Narrative
A ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The non-ge suction assembly was recommended for replacement to resolve the issue.No report of patient involvement.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.(b)(4): a ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The non-ge suction assembly was recommended for replacement to resolve the issue.Block a: no report of patient involvement.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVANCE CS2
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer (Section G)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer Contact
anthony amenson
MDR Report Key15562620
MDR Text Key306077527
Report Number2112667-2022-03983
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeRO
PMA/PMN Number
K172045
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 10/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/12/2022
Initial Date FDA Received10/07/2022
Date Device Manufactured12/03/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-