COVIDIEN MFG SOLUTIONS S.A. DIALYSIS UNKNOWN; CATHETER, PERITONEAL, LONG-TERM INDWELLING
|
Back to Search Results |
|
Model Number DIALYSIS UNKNOWN |
Device Problems
Break (1069); Detachment of Device or Device Component (2907)
|
Patient Problem
Bradycardia (1751)
|
Event Type
Injury
|
Event Description
|
According to the reporter, the clinician recalled a historical event where a hemodialysis device had been in place for over four years and was described as "very old" with deterioration.During assessment of the device for potential interlumen void, the luer came out of the clear rubber portion of the adapter.The physician driven testing procedure inadvertently revealed that there had been a previously unrecognized break/crack/leak in the luer of the catheter's extension tube (unknown if venous/arterial and additional information unavailable at this time).The clinician hypothesized that this had been the source of multiple bradycardic events requiring resuscitation, from presumed air emboli, while undergoing dialysis therapy.Decompensation due to potential air embolism and patient required resuscitation with each event.Once luer deterioration was identified, the catheter was exchanged for a new device (unknown product that will not be made available at this time).The clinician also included that upon device revision/replacement, the d uration of implantation and due to extreme age of the device had caused the catheter to become "stuck" which required an angioplasty procedure to remove the device.Patient condition after resolution of events was stable/return to baseline.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection noted the luer adapter was found to be cracked, leaking, or broken.It was reported that the luer adapter detached from the extension tube.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|