• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ALERE DETERMINE HIV 1/2 AG/AB COMBO 25T; TEST, HIV DETECTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ALERE DETERMINE HIV 1/2 AG/AB COMBO 25T; TEST, HIV DETECTION Back to Search Results
Catalog Number 7D2648
Device Problem False Positive Result (1227)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2022
Event Type  malfunction  
Event Description
The consumer reported two (2) false positive results with determine hiv 1/2 ag/ab combo performed on an an unknown date with an unknown sample.This mfr.Report addresses test results one (1) of two (2).Although requested, no additional patient information, including treatment and outcome, was provided.
 
Manufacturer Narrative
The remainder of the investigation remains in progress.A supplemental report will be provided after completion.Please reference all related mfr.Report numbers: 1221359-2022-04502.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the investigation conclusion.Please see updates: g3, g6, h2 and h6.Testing was performed at abbott diagnostics scarborough, inc.On retained kit lot 177935 with internal positive quality control samples and negative quality control swabs.All test results were valid and performed as expected.Additionally, the manufacturing records and quality control release testing was reviewed for test kit part number 7d2648/ lot 177935 and test base part number 10732998 / lot 169496.The lot met the required release specifications.A review of the complaints reported as false negative patient results (confirmed and unconfirmed, conflicting results) related to kit lot 177935 showed that the complaint rate is (b)(4).Abbott diagnostics scarborough was unable to determine the exact root cause of the reported issue, however it could have possibly been related to the specific patient sample.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALERE DETERMINE HIV 1/2 AG/AB COMBO 25T
Type of Device
TEST, HIV DETECTION
Manufacturer (Section D)
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
10 southgate road
scarborough,
ME 04074
Manufacturer Contact
rachel blackwell
10 southgate road
scarborough,
MDR Report Key15565666
MDR Text Key302771793
Report Number1221359-2022-04501
Device Sequence Number1
Product Code MZF
UDI-Device Identifier10811877010293
UDI-Public011081187701029310177935
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
BP120037-0
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number7D2648
Device Lot Number177935
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/29/2022
Initial Date FDA Received10/07/2022
Supplement Dates Manufacturer Received10/13/2022
Supplement Dates FDA Received10/19/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-