Model Number 0800-0543
Device Problems Erratic or Intermittent Display (1182); No Display/Image (1183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/26/2022
Event Type  malfunction  
Manufacturer Narrative
The customer declined the option to have their glidescope spectrum smart cable returned to verathon for evaluation.Since the device was not returned to verathon for evaluation, the cause could not be determined.Based on communication with the customer, the likely cause of the reported issue was isolated to just the glidescope spectrum smart cable.Review of complaint history for the reported smart cable serial number "(b)(4)" did not identify any previous complaints.Trending analysis for glidescope spectrum smart cables does not identify any trends exceeding acceptable limits.Review of the system risk assessment confirmed that the risk associated with the hazardous scenario has been adequately captured and characterized.Verathon recommended the customer purchase a replacement glidescope spectrum smart cable as their smart cable likely needed to be replaced, and there are currently no repairs available for this device.Corrective action is not required at this time.Verathon will continue to monitor for trends.
 
Event Description
A customer reported that during a patient procedure, using a glidescope spectrum smart cable, the smart cable doesn't connect when the video is needed.It was reported that the video was intermittent when the smart cable was manipulated.No delay during the procedure, use of a backup device, or harm to the patient or user was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESCOPE SPECTRUM SMART CABLE
Type of Device
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA  V5C 5A9
Manufacturer (Section G)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA   V5C 5A9
Manufacturer Contact
corey kasbohm
20001 n creek pkwy
bothell, WA 98011-8218
4256295760
MDR Report Key15580494
MDR Text Key307164664
Report Number9615393-2022-00183
Device Sequence Number1
Product Code CCW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/11/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0800-0543
Device Catalogue Number0800-0544
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/14/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-