• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN DOMINICIAN REPULIC INC. HEMODIALYSIS BLOODLINES; SET, TUBING, BLOOD, WITH A

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN DOMINICIAN REPULIC INC. HEMODIALYSIS BLOODLINES; SET, TUBING, BLOOD, WITH A Back to Search Results
Model Number SL-2010M2096A
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/12/2022
Event Type  malfunction  
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).Two (2) photographs were provided for evaluation.The photographs were visually evaluated, and it was noted one photograph was of the lot# and item #.The other photograph was set full of blood and what appeared to be the arterial being held in operators' hand.The reported of arterial pod separated from bloodline is difficult to determine from the photograph provided.Based on the evaluation the reported defect was unable to be confirmed.Retained units were evaluated and passed the internal testing.A review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.Additionally, review of the batch history records was also performed for the reported lot number and no non-conformances or deviations were noted during the manufacturing process or final inspections.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Event Description
As reported by the user facility: detailed inquiry description - arterial pod separated from the bloodline post treatment when removing bloodlines from the machines.No injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMODIALYSIS BLOODLINES
Type of Device
SET, TUBING, BLOOD, WITH A
Manufacturer (Section D)
B. BRAUN DOMINICIAN REPULIC INC.
las americas industrial park
km22 autopista las americas
santo domingo
DR 
Manufacturer (Section G)
B. BRAUN DOMINICIAN REPULIC INC.
las americas industrial park
km22 autopista las americas
santo domingo
DR  
Manufacturer Contact
jonathan severino
901 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key15581004
MDR Text Key302730654
Report Number2523676-2022-00502
Device Sequence Number1
Product Code FJK
UDI-Device Identifier04046955674992
UDI-Public(01)04046955674992
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K080807
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSL-2010M2096A
Device Catalogue NumberSL-2010M2096A
Device Lot Number0061804217
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/14/2022
Initial Date FDA Received10/11/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/22/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-