This report has been identified as b.Braun medical internal report number (b)(4).Two (2) photographs were provided for evaluation.The photographs were visually evaluated, and it was noted one photograph was of the lot# and item #.The other photograph was set full of blood and what appeared to be the arterial being held in operators' hand.The reported of arterial pod separated from bloodline is difficult to determine from the photograph provided.Based on the evaluation the reported defect was unable to be confirmed.Retained units were evaluated and passed the internal testing.A review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.Additionally, review of the batch history records was also performed for the reported lot number and no non-conformances or deviations were noted during the manufacturing process or final inspections.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|