Model Number PE075F5 |
Device Problem
Pacing Problem (1439)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/20/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device evaluation is anticipated.However, the complaint cannot be confirmed without the completion of a product evaluation.A supplemental report will be forthcoming with the evaluation.A device history record review was completed and documented that device met all specifications upon distribution.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
|
|
Event Description
|
It was reported that no continuity in the circuit was confirmed during continuity test before the swan ganz catheter insertion.The issue was resolved by replacing the catheter.Information such as what kind of surgery or examination the catheter was used for or if the patient had cardiac conduction defect is unknown.Patient demographic information was requested but unavailable.There were no patient complications reported.
|
|
Manufacturer Narrative
|
The reported event of no continuity in the circuit was confirmed.Continuity testing found that a short condition occurred between the proximal and distal circuits in the y adaptor.No open or short condition was observed in the lead wires between distal side of y adaptor and the electrodes.No visible damage was observed from catheter body, balloon, and windings.The balloon inflated clear and concentric and remained inflated for 5min without leakage.Further evaluation regarding related quality issues is under investigation.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
|
|
Manufacturer Narrative
|
Engineering evaluation was completed and concluded that a product risk assessment escalation was met earlier.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
|
|