• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER INFINITI VISION SYSTEM, ACCESSORY, OZIL TORSIONAL HANDPIECE; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER INFINITI VISION SYSTEM, ACCESSORY, OZIL TORSIONAL HANDPIECE; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065750469
Device Problem Detachment of Device or Device Component (2907)
Patient Problems Capsular Bag Tear (2639); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 07/28/2022
Event Type  Injury  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A nurse reported that when patient was undergoing a phacoemulsification with intraocular lens implant for left eye cataract removal.During a critical point in the procedure the tubing disconnected from the ophthalmic consoles handset and a sudden shallowing of the anterior chamber occurred.The equipment was inspected and tested to be working again the surgeon noted a posterior capsule rupture.An anterior vitrectomy was then preformed.
 
Manufacturer Narrative
The product under investigation is not a serviceable device.Therefore, a service record review was not performed.There was no equipment returned for this investigation.Based on the information obtained, the root cause of the reported event is inconclusive.A non-conformance based review of the batch/lot/serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this lot/batch/serial number was performed.No similar complaints were reported for the product lot/batch/serial under investigation.Based on the information obtained, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITI VISION SYSTEM, ACCESSORY, OZIL TORSIONAL HANDPIECE
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15593033
MDR Text Key301630057
Report Number2028159-2022-01450
Device Sequence Number1
Product Code HQC
UDI-Device Identifier00380657504695
UDI-Public00380657504695
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112425
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 12/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/13/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number8065750469
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/22/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/06/2013
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
CENTURION SURGICAL PROCEDURE PAK
Patient Outcome(s) Other; Required Intervention;
Patient Age74 YR
Patient Weight110 KG
-
-