(b)(6), from thrive skilled pediatric care stated, "they were called by the hospital, told the patient had expired an asked if they could come pick up the vent, they never said anything about the vent malfunctioning.".
|
The device was returned, and we found no faults with this device.We looked at the alarm/events log and saw a lot of high/low pressure alarms, which could be caused by faulty circuits or incorrect settings selection.The device passed functional testing.There is a decent sized crack on the top screen.This is purely cosmetic damage.The conclusion is that the device operated as intended.Due to the time lapse between the adverse event (1 jan 2022) and this analysis, it cannot be concluded whether the crack was present during the event.Since it is cosmetic in nature, it would not have contributed to the event.Analysis of log files is not possible.The log files pertaining to the adverse event had been erased and it is therefore not possible to analyze any log files for this event.The conclusion of the investigation is that the device performs as intended and according to specification, and no correlation between the reported event and any device malfunction can be established.
|