• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC AUTOCAT2 WAVE; SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC AUTOCAT2 WAVE; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Model Number IPN000320
Device Problem Failure to Run on Battery (1466)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/15/2022
Event Type  malfunction  
Manufacturer Narrative
Qn# (b)(4).Other remarks: n/a.Corrected data: n/a.
 
Event Description
As reported "the issue occurred during incoming inspection/inventory review, product training/demonstration." additional information states "pump will not run on battery, will only run with ac power when battery breaker is tripped.Also when attempting to pump, erratic bpw and multiple purge, high baseline alarms." no report of patient involvement.
 
Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.Additionally no recorder strip was returned for investigation.Per service database, the iabp has not had performed servicing since 29 sep 2022.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed, and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.Other remarks: n/a.Corrected data: n/a.
 
Event Description
As reported "the issue occurred during incoming inspection/inventory review, product training/demonstration." additional information states "pump will not run on battery, will only run with ac power when battery breaker is tripped.Also when attempting to pump, erratic bpw and multiple purge, high baseline alarms." no report of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOCAT2 WAVE
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
bryanna connelly
3015 carrington mill blvd
morrisville 27560
MDR Report Key15594574
MDR Text Key304624011
Report Number3010532612-2022-00403
Device Sequence Number1
Product Code DSP
UDI-Device Identifier30801902051715
UDI-Public30801902051715
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K060309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberIPN000320
Device Catalogue NumberIAP-0500
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/15/2022
Initial Date FDA Received10/13/2022
Supplement Dates Manufacturer Received11/04/2022
Supplement Dates FDA Received11/04/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-