Catalog Number C49291349-NLJ
Device Problem Gas/Air Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/01/2021
Event Type  malfunction  
Manufacturer Narrative
A sample was received to perform an investigation.A visual inspection of the returned device found a device free of damage.No loose circuit was observed in the sample.Functional testing was performed using a leak test.The device passed the leak test.The reported problem was unable to be replicated.No root cause could be determined as the complaint could not be confirmed.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.Udi, operator of device and pma/510(k) number are unknown.No information has been provided to date.This remediation mdr was generated under protocol b10009406, as a result of warning letter cms# (b)(4).
 
Event Description
It was reported that during a hospital pre-use check, it was noticed that the connector of the breathing circuit was loose, which caused an air leak to occur.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX ANESTHESIA BREATHING CIRCUITS
Type of Device
CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
2-12-12 kitanagase-omotemachi,
minneapolis, MN 55442
MDR Report Key15595880
MDR Text Key306520251
Report Number3012307300-2022-20057
Device Sequence Number1
Product Code CAI
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date06/01/2023
Device Catalogue NumberC49291349-NLJ
Device Lot Number3997535
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/08/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/10/2022
Initial Date FDA Received10/13/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/02/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-