ABBOTT VASCULAR RX ACCULINK CAROTID STENT SYSTEM Back to Search Results
Catalog Number 1010132-40
Device Problems Malposition of Device (2616); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/09/2022
Event Type  malfunction  
Manufacturer Narrative
The stent remains in the patient.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that the procedure was to treat the right carotid artery with heavy calcification and heavy tortuosity.The lesion was crossed with an 0.014 guide wire and then the 6-8x40 mm acculink self expanding stent system (sess) was advanced to the lesion.During deployment there was resistance with the handle of the device.The stent was deployed; however, the stent jumped on deployment.A 1/3 of the stent remains in the target lesion.To cover the rest of the target lesion another acculink stent was deployed well without any resistance.A viatrac balloon dilatation catheter (bdc) was used to post dilate and complete the procedure.There were no adverse patient effects and there was no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.As there was no damage noted to the device during the inspection prior to use, the investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that during deployment interaction with the heavily calcified and heavily torturous anatomy resulted in preventing the shaft lumens from moving freely; thus resulting in the reported physical resistance (resistance with the handle).Interaction with the heavily calcified, heavily torturous anatomy and/or manipulation of the device during stent deployment resulted in the stent to jump on deployment (reported stent malposition).There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX ACCULINK CAROTID STENT SYSTEM
Type of Device
CAROTID STENT
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15596849
MDR Text Key304011914
Report Number2024168-2022-10616
Device Sequence Number1
Product Code NIM
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
P040012
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2024
Device Catalogue Number1010132-40
Device Lot Number2040861
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/27/2022
Initial Date FDA Received10/13/2022
Supplement Dates Manufacturer Received10/25/2022
Supplement Dates FDA Received11/15/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/08/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
0.014 GUIDE WIRE
-
-