Model Number N/A |
Device Problems
Thermal Decomposition of Device (1071); Melted (1385); Failure to Power Up (1476); Smoking (1585)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/23/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that the power cord was melted, and the unit no longer powers on.Upon follow up it was noted that the unit displayed charring and there was cord damage.No adverse event was reported as it relates to the event.This event took place outside of surgery and there was no patient involvement.Due diligence is complete and no additional information is available.
|
|
Manufacturer Narrative
|
This event has been reported under (b)(4).Upon completion of investigation a final/supplemental report will be submitted.
|
|
Event Description
|
No additional event information available.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.Review of the most recent repair record determined that the power inlet module fuse was tripped and the power inlet module and power cord were damaged and charred.The power inlet module and power cord were replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.The root cause of the reported issue is attributed to the time the high flow valve is left open being too short during the pump start up sequence.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Search Alerts/Recalls
|