Catalog Number 1014261-080
Device Problems Inflation Problem (1310); Stretched (1601)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/09/2022
Event Type  malfunction  
Manufacturer Narrative
Visual and functional inspections were performed on the returned device.The reported inflation issue was not confirmed.Additionally, return analysis noted the outer member was stretched, proximal to the proximal balloon seal, possibly due to inadvertent mishandling of the device; however, it is unknown when this damage occurred.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported inflation issue.There is no indication of a product quality issue with respect to design, manufacture or labeling of the device.
 
Event Description
It was reported that the procedure was to treat a moderately calcified, mildly tortuous left anterior tibial artery.The 3.0x80mm armada 18 balloon dilatation catheter (bdc) did not fully expand when attempting to inflate.Another unspecified device was used to successfully complete the procedure.There was no adverse patient effects reported and no clinically significant delay reported.Per device analysis the outer member was stretched, 3mm proximal to the proximal balloon seal for a length of 2mm.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 18 PTA CATHETER
Type of Device
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15607670
MDR Text Key306988962
Report Number2024168-2022-10651
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K151317
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 10/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2023
Device Catalogue Number1014261-080
Device Lot Number0030541
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/22/2022
Initial Date FDA Received10/14/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/05/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-