• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM, ACCESSORY, ACTIVE SENTRY HANDPIECE; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM, ACCESSORY, ACTIVE SENTRY HANDPIECE; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065752914
Device Problems Overheating of Device (1437); Device Displays Incorrect Message (2591); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported that an ophthalmic handpiece gets overheated and would not cool before surgery.The procedure details and patient impact were not reported.Additional information was received and indicated that handpiece would not engage and the surgery was completed with another handpiece.There was no patient involvement.
 
Manufacturer Narrative
Removed fda code 2907 and 2591.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
The product under investigation is not a serviceable device.Therefore, a service record review was not performed.However, the hp (handpiece) was returned for this investigation.A non-conformance-based review of the batch/lot/serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this lot/batch/serial number was performed.No similar complaints were reported for the product lot/batch/serial under investigation.The hp was received for this investigation.A visual assessment of the returned sample found that no nonconformities were observed.The sample was connected to a calibrated resistance breakout box, where its input and/or output impedance was found to be out of specification.The returned sample was connected to a calibrated system, where it displayed sm, which is indicative of a nonconforming handpiece pressure sensor.Due to the condition of the hp, further testing could not be performed the root cause of the reported event system message (sm) can be attributed to wear/reliability of the sensor pressure.However, how or when this component became nonconforming remains inconclusive.Based on the information obtained, the root cause of the high temperature is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM, ACCESSORY, ACTIVE SENTRY HANDPIECE
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15612061
MDR Text Key306429874
Report Number2028159-2022-01457
Device Sequence Number1
Product Code HQC
UDI-Device Identifier00380657529148
UDI-Public00380657529148
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161794
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 01/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/14/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065752914
Device Lot Number146NYX
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/28/2022
Date Manufacturer Received01/03/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/10/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-