STRYKER ENDOSCOPY-SAN JOSE SERFAS ENERGY 90-S CRUISE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 0279401200
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Patient Device Interaction Problem (4001)
Patient Problem Burn(s) (1757)
Event Date 09/12/2022
Event Type  Injury  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event Description
It was reported that the patient was burned.
 
Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: patient burned probable root cause: heat can result from insufficient suction which can result from: * inadequate hospital suction, leak, bad connection to hospital suction line, * clogging caused by suction path blockage, lower electrode mass due to variation in electrode thickness or opening cut use error the reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.H3 other text : 81.
 
Event Description
It was reported that the patient was burned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SERFAS ENERGY 90-S CRUISE
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
andrea zenere
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key15620621
MDR Text Key301895102
Report Number0002936485-2022-00590
Device Sequence Number1
Product Code GEI
UDI-Device Identifier07613327057034
UDI-Public07613327057034
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K160050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0279401200
Device Catalogue Number0279401200
Device Lot NumberNR
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/21/2022
Initial Date FDA Received10/17/2022
Supplement Dates Manufacturer Received09/21/2022
Supplement Dates FDA Received11/28/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-