Model Number N/A |
Device Problems
Thermal Decomposition of Device (1071); Failure to Power Up (1476); Suction Problem (2170)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/01/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event has been recorded under (b)(4).Once investigation is completed a final/supplemental report will be submitted.
|
|
Event Description
|
It was reported outside of surgery, the device had no suction which was a result of a burnt power inlet module.No patient involvement was noted as a result no harm or delay occurred.No adverse event was reported as it relates to this event.Diligence is completed and no additional information is available.
|
|
Manufacturer Narrative
|
Review of the most recent repair record determined that the power inlet module was burnt and the unit was not powering on.The power inlet module was replaced and resolved the reported issue.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
There is no additional event information available.
|
|