EDWARDS LIFESCIENCES LLC; CATHETER, UMBILICAL ARTERY Back to Search Results
Lot Number 64135772
Device Problem Defective Device (2588)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/23/2022
Event Type  malfunction  
Event Description
Line change was due for umbilical artery catheter (uac) fluids, the line was changed and hung at 1:20 am.Blood pressures were appropriate and line seemed to be working fine.Then blood pressure means were 66/63 and then not reading.The line was assessed and flushed and seemed to be better.The diaper cloth under the line was noted to be wet but multiple registered nurses (rns) could not figure out where the wet feeling was coming from.Line flushed again and no leaking to be noted.Line checked again at 0400 because it was not working again and leaking noted from the tubing (not at a luer lock connection site).Registered nurse called pharmacy and ordered new uac fluids.Line tubing was then changed and malfunctioning tubing saved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
CATHETER, UMBILICAL ARTERY
Manufacturer (Section D)
EDWARDS LIFESCIENCES LLC
one edwards way
irvine CA 92614
MDR Report Key15629324
MDR Text Key301998910
Report Number15629324
Device Sequence Number1
Product Code FOS
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Risk Manager
Type of Report Initial
Report Date 09/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Lot Number64135772
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA09/27/2022
Event Location Hospital
Date Report to Manufacturer10/18/2022
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/18/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age10 DA
Patient SexMale
-
-