Model Number 456020 |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/13/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
This report has been identified as b.Braun medical internal report number: (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
|
|
Event Description
|
As reported by the user facility: medication tubing and stopcock changed @ 00:01h (for propofol infusion) on (b)(6) 2022.Connection patent/leak free at that time.@ 0110h - rn noted pooling of opaque medication on bed (propofol).This rn noted opaque medication beyond stopcock still infusing (difficult to read in document), as well as pooled under connection.Infusion set cracked at hub of set/stopcock connection.
|
|
Manufacturer Narrative
|
This report has been identified as b.Braun medical internal report number (b)(4).No sample was provided for evaluation.Based on the data from the investigation we are unable to determine the root cause of the reported incident.The reported defect was unable to be confirmed.The actual defective device is a valuable tool in investigating the cause of this incident.Retained units were evaluated and passed the internal testing.Review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|
|