• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MEDICAL INC. DISCOFIX®; ACCESSORIES, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC. DISCOFIX®; ACCESSORIES, CATHETER Back to Search Results
Model Number 456020
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/16/2022
Event Type  malfunction  
Event Description
As reported by the user facility: description: stopcocks (blue) leaking despite tightening and securing all stopcocks.Leaking sedation medications, all stopcocks changed but still leaking.Old stopcocks (yellow) in stock on unit so blue stopcocks all thrown away from patient and changed to yellow stopcocks.No injury reported.
 
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).One (1) used sample with packaging was returned for evaluation.The sample was visually and physically tested as per specification with passing results.No leakage was observed on the discofix.Based on the evaluation results, the reported defect was not confirmed.Retained units were evaluated and passed the internal testing.Review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISCOFIX®
Type of Device
ACCESSORIES, CATHETER
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd.
allentown PA 18109
Manufacturer Contact
jonathan severino
901 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key15632699
MDR Text Key302015550
Report Number2523676-2022-00512
Device Sequence Number1
Product Code KGZ
UDI-Device Identifier04046964996962
UDI-Public(01)04046964996962
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K760383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number456020
Device Catalogue Number456020
Device Lot Number21M11D9040
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/04/2022
Initial Date FDA Received10/19/2022
Supplement Dates Manufacturer Received10/04/2022
Supplement Dates FDA Received11/29/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/18/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-