ZIMMER BIOMET, INC. G7 CURVED INSERTER THD SHAFT; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/29/2022
Event Type  malfunction  
Event Description
It was reported when the surgeon was impacting the cup, the threads were constantly loosing.We tried 4 different threaded shafts before we found one that holds the cup property.No additional information.
 
Manufacturer Narrative
(b)(4).Foreign country: canada.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2022 - 02364, 0001825034 - 2022 - 02366, 0001825034 - 2022 - 02367.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Lot identification is necessary for review of device history records, lot identification was not provided.Devices are used for treatment.Review of complaint history identified additional similar complaints for the reported item(s).However, as the lot number is unknown, an additional review could not be performed.Medical records were not provided.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
There is no update to the original complaint description provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 CURVED INSERTER THD SHAFT
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15633921
MDR Text Key302363526
Report Number0001825034-2022-02365
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/19/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number110003454
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-