This follow-up is being submitted to relay additional information.Updated: b4, b5, d4, g3, g6, h1, h2, h3, h4, h6, and h10 the event is confirmed.Review of the most recent repair record determined the unit was confirmed to not power on.The power inlet module had melted slightly and there were no signs of burning smoke or charring.The power inlet module was replaced and resolved the reported issue.The root cause of the reported issue is attributed to the time the high flow valve is left open is too short during the pump start up sequence.If any further information is which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|