STERIS CORPORATION RELIANCE; DISINFECTOR, MEDICAL DEVICES Back to Search Results
Model Number 1227
Device Problem Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2021
Event Type  malfunction  
Event Description
Reliance® 1227 cart and utensil washer/disinfector displayed alarm #7, recirculation error, on the screen for staff, health technology management responded and worked with steris to resolve the issue.Over the course of three months, the replacement of all parts of the system affected, multiple visits from steris, it was determined to be a door sensor that failed.The general nature of the error with no way of seeing what generated it in the system from either the hospital or steris technicians resulted in excessive cost of repair and an unreasonable down time for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RELIANCE
Type of Device
DISINFECTOR, MEDICAL DEVICES
Manufacturer (Section D)
STERIS CORPORATION
5960 heilsey road
mentor OH 44060
MDR Report Key15635740
MDR Text Key302067318
Report Number15635740
Device Sequence Number1
Product Code MEC
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/28/2022,09/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number1227
Device Catalogue NumberFW03
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA09/28/2022
Event Location Hospital
Date Report to Manufacturer10/19/2022
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/19/2022
Type of Device Usage Unknown
Patient Sequence Number1
-
-