Model Number CI-1400-01 |
Device Problems
Mechanical Problem (1384); Battery Problem (2885); Intermittent Communication Failure (4038)
|
Patient Problems
Insufficient Information (4580); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 10/24/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Event Description
|
The recipient is reportedly experiencing low battery life and intermittent lock.Device testing revealed results within normal limits.External equipment was exchanged, however, the issue did not resolve.Revision surgery is scheduled.
|
|
Manufacturer Narrative
|
The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed sliced silicone overmold on the top and bottom covers, as well as a severed electrode.These anomalies are believed to have occurred during revision surgery.The device passed photographic imaging inspection.System lock could not be obtained at any spacing.The electrode condition prevented an electrical test from being performed.The device passed some of the electrical tests performed.The device failed the residual gas analysis test.This device had moisture that exceeded the residual gas analysis test limit.Based on the assessment of the residual gas analysis data, it is determined that this device was non-hermetic.Corrective actions were implemented.This version of the 90k device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|