• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OAKDALE LEVEL 1 HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OAKDALE LEVEL 1 HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-90
Device Problem Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).The original device history record (dhr) is no longer applicable as device was manufactured in 2004.The result of the investigation do not implicate a service process.The device was received for evaluation; device was received with some cracks in the enclosure.The device was manufactured in 2004.Functional test: filled the reservoir tank with water to check for leaks.Device powered on to see if the lcd would turn on.At this point, the pump had powered on, however the lcd was blank, with no display.This confirmed customer reported problem.Blank lcd display was caused by the electronics assembly not working as intended.Device was 15 years old and main board was the old version.Device also had design issue, which was addressed through a corrective and preventative action.Root cause of the issue was deemed to be normal wear and tear.Device was 15 years old.Action(s) taken to mitigate the reported issue(s): replaced the pcb (printed circuit board).
 
Event Description
It was reported that the lcd did not turn on.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15639533
MDR Text Key306889375
Report Number3012307300-2022-25402
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier50695085820909
UDI-Public50695085820909
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/20/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-90
Device Catalogue NumberHL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/18/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/26/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/30/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-